USING THE ABBE APERTOMETER

The light path when using the apertometer is shown in the images. Light is directed horizontally towards the curved side, and is reflected by the 45 degree angle at the back of the glass upwards through the silver disc opening. The light source may need to be repositioned as the cursor is moved. Using the Abbe apertometer to determine NA is usually reasonably accurate, provided instructions for its use are carefully followed. The instructions by Zeiss leave out a few helpful details on improving accuracy mentioned by Spitta and in greater detail by Hartridge. Hartridge goes further to make several suggestions for improving accuracy employing a reversed light path, that is with the light directed down the optical tube and the NA determined by the pattern of light falling on the apertometer. In addition, several other caveats about Hartridge's method make his method quite tedious. That degree of accuracy, though it may be improved with his method, is not neccesary for ordinary work, as the errors occurring when the device is employed when the device is used as originally intended, are too small to be detected by the human eye. That is, the human eye cannot differentiate a change in resolution with small changes in NA. A satisfactory result can be obtained by combining the instructions by Zeiss and the precautions mentioned by Spitta.
When I tested some modern objectives, the results were as follows:
In use, testing just three objectives, I found the following: a 10X S-Plan olympus objective labeled as NA of 0.30, measured an NA of 0.35. A generic 100X objective, labeled as NA of 1.3 had a measured NA using the Abbe apertometer of 1.23. Finally an apochromatic Olympus S-plan objective labeled as NA of 1.35 measured 1.30.
Outline of use:
- Place the instrument on the stage
- Center the little coverslip under the objective
- With a bright light source aimed a the Apertometer's round side, more precisely center the line between the two hemicircles of silver under the coverslip and focus on the edge of these silver hemicircles, using the usual method(left)
- Taking care not to disturb the focus, attach the special objective to the inside distal end of the drawtube and slide it in or out to focus on the cross of the cursor. Alternatively, use a phase telescope or Bertrand lens to do the same. This will obviate the need for using the auxilliary objective (right).
- Move the cursor to one side and note the reading where the center of the X reaches the edge of the field; do the same in the other direction
- The NA is the average of the two readings
- Note that the above can be performed with the naked eye alone with the eyepiece removed for the lower power objectives. An auxilliary magnification system is not needed for low power objectives.
- Note that immersion oil must be used for oil immersion objectives.
- Note that with the older model with two indicators, one has to focus on the edge of the indicator as opposed to the cross in the cursor.
VIEW OF THE EDGE OF THE SLIT UNDER THE MICROSCOPE
VIEW OF THE X IN THE CURSOR AS VIEWED THROUGH A PHASE TELESCOPE OR BERTRAND LENS AS IT APPROACHES THE EDGE OF THE FIELD
One annoying problem I noticed in using this apertometer. The instrument tends to move when the cursor is being moved; a mechanical stage clip is not strong enough to prevent slipping.
Pushing down on the apertometer to hold in place by hand will also lower the stage; to prevent this, a locking mechanism for the coarse focusing control is helpful. I note that in some examples, the Abbe apertometer came with pins to fit in the holes on the stage for the stage clips to prevent this annoyance. This would not be practical unless the pins were made for a specific single microscope, as most microscopes have the holes for the stage clips different distances apart and these holes are found in more than one diameter.